Fluke CD Player 6000D User Manual

Manual Supplement  
Manual Title:  
Print Date:  
Revision/Date: 1, 1/08  
Impulse 6000D/7000DP Users Supplement Issue:  
3
3/09  
10  
August 2007  
Issue Date:  
Page Count:  
This supplement contains information necessary to ensure the accuracy of the  
above manual. This manual is distributed as an electronic manual on the  
following CD-ROM:  
CD Title:  
CD Rev. & Date:  
CD PN:  
6000D/7000DP  
1, 2/2008  
3028681  
© 2008-2009 Fluke Corporation. All rights reserved. Printed in the U.S.A.  
 
Manual Supplement  
Impulse 6000D/7000DP Users  
Accuracy............................................................. ±0.1 ms  
Voltage  
Range ................................................................. 20 to 5000 V  
Accuracy............................................................. ±(1 % of reading + 2 V)  
Current  
Range ................................................................. 0.4 to 100.0 A  
Accuracy............................................................. ±(1 % of reading + 0.1 A)  
Tilt (biphasic and pulsed biphasic)  
Range................................................................. 1 % to 99 %  
Accuracy............................................................. ±1 digit  
Interphase delay (biphasic and pulsed biphasic)  
Range................................................................. 0.1 ms to 9.9 ms  
Accuracy............................................................. ±0.1 ms  
Frequency (pulsed biphasic only)  
Range................................................................. 2000 Hz to 8000 Hz  
Accuracy............................................................. ±1 % of reading  
Duty cycle (pulse biphasic only)  
Range................................................................. 1 % to 99 %  
Accuracy............................................................. ±1 digit  
Sample rate ........................................................... 250 kHz (4 μs sample)  
Maximum Average Power.................................... 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes  
2
8/08  
 
Impulse 6000D/7000DP Users  
Manual Supplement  
Scope Output  
Autorange ...........................................................2000:1, 400:1 and 80:1: dependant on the range  
Waveform Playback  
Output .............................................................BNC  
Output impedance...........................................50 Ω  
Amplitude Accuracy ........................................±5 %  
Charge Time Measurement  
Range .................................................................0.1 to 100.0 s  
Accuracy .............................................................±0.05 s, typical  
Synchronization Test (Elective Cardioversion)  
Delay Time Measurement  
Timing window ................................................ECG R-wave peak to the defib pulse peak  
Range .............................................................-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380  
ms following the R-wave peak.  
Resolution.......................................................1 ms  
Accuracy .........................................................±1 ms  
ECG waves  
Normal Sinus Rhythm (NSR)..........................10 to 180 (by 1) BPM  
Atrial fibrillation................................................Coarse and fine  
Monomorphic Ventricular Tachycardia ...........120 to 240 (by 5) BPM  
Asystole ..........................................................Flat line  
Automated Defibrillator Test ECG Waves  
Normal Sinus ......................................................10 to 300 (by 1) BPM  
4/08  
3
 
Manual Supplement  
Impulse 6000D/7000DP Users  
Ventricular Fibrillation......................................... Coarse and fine  
Monomorphic Ventricular Tachycardia............... 120 to 300 (by 5) BPM  
Polymorphic Ventricular Tachycardia................. 5 types  
Asystole .............................................................. Flat line  
ECG Waves  
ECG General  
Lead configuration.............................................. 12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs  
Lead to lead impedance ..................................... 1000 (nominal)  
Rate accuracy..................................................... ±1 % of nominal  
ECG Amplitudes  
Reference Lead...................................................Selectable, Lead II (default) or Lead I  
Settings................................................................0.05 to 0.45 (by 0.05) mV  
0.5 to 5.0 (by 0.5) mV  
Accuracy (All Performance waves and Normal Sinus R waves)  
Lead II..................................................................±2 %  
All other leads......................................................±5 %  
Defib paddles.......................................................±5 %  
Amplitude of ECG signals relative to amplitude setting (in percent)  
Lead II reference  
Performance waves and R wave detection  
Lead #  
I
II  
III  
V1  
V2  
V3  
V4  
V5  
V6  
Ref. Amp.  
70 %  
100 %  
30 %  
100 %  
100 %  
100 %  
100 %  
100 %  
100 %  
4
3/09  
 
Impulse 6000D/7000DP Users  
Manual Supplement  
Normal Sinus waves  
Lead #  
I
II  
III  
V1  
V2  
V3  
V4  
V5  
V6  
Ref. Amp.  
70 %  
100 %  
30 %  
24 %  
48 %  
100 %  
120 %  
112 %  
80 %  
Lead I reference  
Performance waves and R wave detection  
Lead #  
Ref. Amp.  
I
II  
III  
V1  
V2  
V3  
V4  
V5  
V6  
100 %  
150 %  
50 %  
100 %  
100 %  
100 %  
100 %  
100 %  
100 %  
Normal Sinus waves  
Lead #  
I
II  
III  
V1  
V2  
V3  
V4  
V5  
V6  
Ref. Amp.  
100 %  
150 %  
50 %  
24 %  
48 %  
100 %  
120 %  
112 %  
80 %  
ECG Normal Sinus  
Rates...................................................................10 to 360 (by 1) BPM  
ECG High Level Output (BNC Jack)  
Amplitude  
Range .............................................................0.5 V per mV of reference lead setting  
Accuracy .........................................................±5 %  
Output Impedance ..............................................50 Ω  
ECG on Defibrillator Input Load  
Same as the LEAD II amplitude but limited to ±4 mV  
ECG Performance Waves  
Square wave.......................................................2.0 and 0.125 Hz  
Triangular wave ..................................................2.0 and 2.5 Hz  
8/08  
5
 
Manual Supplement  
Impulse 6000D/7000DP Users  
Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz  
Pulse................................................................... 30 and 60 BPM, 60 ms pulse width  
R-Wave Detection  
Waveform ........................................................... Haver-triangle  
Amplitude............................................................ 0.05 to 0.45 (by 0.05) mV  
0.5 to 5.0 (by 0.5) mV  
Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM  
Widths................................................................. 8, 10, 12 ms, and 20 to 200 (by 10) ms  
Accuracy............................................................. ±(1 % setting + 1 ms)  
Noise Immunity  
Wave................................................................... Sine  
Line Frequency................................................... 50 or 60 Hz (± 0.5 Hz)  
Amplitude  
Range............................................................. 0.0 to 10.0 (by 0.5) mV  
Accuracy......................................................... ±5 %  
Transvenous Pacer Pulse Simulation  
Widths  
Range............................................................. 0.1, 0.2, 0.5, 1.0, and 2.0 ms  
Accuracy......................................................... ±5 % of setting  
Amplitude  
Range............................................................. 0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and  
±700 mV  
6
4/08  
 
Impulse 6000D/7000DP Users  
Manual Supplement  
Accuracy .........................................................±(10 % of setting + 0.2 mV)  
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)  
Lead II reference  
Lead #  
Ref. Amp.  
I
II  
III  
V1  
V2  
V3  
V4  
V5  
V6  
67 %  
100 %  
33 %  
67 %  
67 %  
67 %  
67 %  
67 %  
67 %  
Lead I reference  
Lead #  
I
II  
III  
V1  
V2  
V3  
V4  
V5  
V6  
Ref. Amp.  
100 %  
150 %  
50 %  
100 % 100 % 100 % 100 % 100 % 100 %  
Arrhythmia Selections  
Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)  
Demand ..........................................................30 to 360 (by 1) BPM  
Asynchronous  
Non-Capture  
Non-Function  
Threshold (Interactive pacing simulation only)10 to 250 (by 10) mA  
Supraventricular  
Atrial Fibrillation Coarse  
Atrial Fibrillation fine  
Atrial Flutter  
Sinus Arrhythmia  
Missed Beat  
Atrial Tachycardia  
4/08  
7
 
Manual Supplement  
Impulse 6000D/7000DP Users  
Change #2  
On page 12, prior to Analyzing Pacemakers (7000DP only), add the following section:  
SCOPE OUTPUT  
Each time a defibrillator is fired into the Analyzer during an active defibrillator test, a delayed representation of the  
defibrillator pulse is sent to the Scope Output jack on the rear panel of the Analyzer. The Scope Output is an  
isolated signal and should not present any problems when connected to the input of an oscilloscope.  
To assist with the displaying of the defibrillator pulse, synchronizing marker pulses appear on the scope output jack  
along with the defibrillator pulse. Table 3A list these markers and describes their purpose.  
Table 3A. Synchronizing Marker Pulses  
Marker  
Defib fire  
Ranging  
Pulse Characteristics  
+2 V, 50 ms  
Purpose  
Indicates when the defibrillator was fired.  
Indicates the scaling used to output the pulse image:  
-4 V, 0.4 ms, 1 ms apart  
1 pulse……….each volt out equals 80 volts pulse amplitude.  
2 pulses……...each volt out equals 400 volts pulse amplitude.  
3 pulses……...each volt out equals 2000 volts pulse amplitude.  
8
4/08  
 
Impulse 6000D/7000DP Users  
Manual Supplement  
Pulse  
+4 V, 1 ms pulse  
Start of defibrillator pulse playback [1]  
playback  
[1] Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early  
transitional signals.  
Use the synchronizing marker pulses to display specific information from the Scope Output signal. To see when the  
defibrillator fires, set the oscilloscope trigger level to +2 V. To get the ranging information, set the oscilloscope to trigger  
on the -4 V markers. To ignore the ranging information and see only the defibrillator pulse, set the oscilloscope to trigger  
on levels above +2 volts.  
On page 26, prior to Controlling the Analyzer Remotely, add the following section:  
Setting the ECG Reference Lead  
ECG signal amplitude is set by designating either lead I or lead II as the reference lead. Lead II is the default  
selection. To change the reference lead, do the following:  
Press Θ to reveal the main setup menu. Next press the softkey labeled More for additional setup  
selections. Press the softkey labeled ECG Ref Lead to toggle between lead I and lead II. Once the  
reference lead is set, press the softkey labeled Back or one of the other functions to store the selection  
in non-volatile memory.  
4/08  
9
 
Manual Supplement  
Impulse 6000D/7000DP Users  
Change #3, 48038  
On page 38, under Pacemaker Input, change:  
From: Accuracy …………..±1 % non-inductive (<2 μH)  
To:  
Accuracy……………±2 % non-inductive (<2 μH)  
10  
8/08  
 

Echo Chainsaw CS 330T User Manual
EdgeStar Beverage Dispenser TBC50S User Manual
Eureka Vacuum Cleaner 400 409 User Manual
Fellowes Paper Shredder DS 1400C User Manual
Fluke Thermometer 1504 User Manual
Fujita Cameras TV DVD Combo CGTV510651 User Manual
Fujitsu Switch FS 1004EX User Manual
Garland Griddle RGM 2424 1 User Manual
GE Convection Oven Convection Grill Combination Microwave Cooking Manual User Manual
Goodman Mfg Heat Pump GSH13 Remote Heat Pump User Manual