ABB CL3020 User Manual

ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION  
CL3020  
CLD NOx analyzer  
Measurement made easy  
2
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Table of Contents  
Important precautions..............................................................................................5  
Use dry, oil-free instrument air only.................................................................5  
Safety notice.........................................................................................................5  
Model CL3020 specifications...................................................................................6  
Performance specifications...............................................................................6  
Features ................................................................................................................6  
Mechanical specifications.................................................................................. 7  
Overview......................................................................................................................8  
Measurement configurations............................................................................8  
Theory of operation...................................................................................................9  
Chemiluminescence measurement of NOx......................................................9  
Zirconium oxide measurement of O2 ...............................................................9  
Pneumatic design..............................................................................................10  
Analyzer setup and quick start procedure........................................................... 13  
Connect the analyzer......................................................................................... 13  
Apply power to the analyzer and check diagnostics ...................................14  
Set analog outputs of the analyzer ................................................................ 15  
Calibrate the analyzer .......................................................................................16  
Display screens and details of operation ............................................................ 17  
Home screen, warnings, and alarms .............................................................. 17  
Calibration menu ...............................................................................................18  
Diagnostic screen..............................................................................................19  
Trend screen.......................................................................................................20  
Config screen .....................................................................................................20  
Analog output scaling screen.......................................................................... 21  
Analog output trim screen...............................................................................22  
IP address screen...............................................................................................23  
QR code ...............................................................................................................23  
About screen ......................................................................................................24  
Alarm screen.......................................................................................................24  
Config alarms screen ........................................................................................25  
Troubleshooting ......................................................................................................26  
Diagnostics and operating parameters ........................................................26  
Warnings and alarms ..................................................................................26  
Sample and ozone flow ..............................................................................26  
Converter temperature ..............................................................................27  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
3
Gas concentration readings ......................................................................27  
System checks....................................................................................................27  
Electrical connections and fuse................................................................27  
Leak checking...............................................................................................28  
Component information ..................................................................................28  
Oxygen sensor module...............................................................................28  
Ozonator.......................................................................................................28  
NOx reaction cell ..........................................................................................28  
Converter ......................................................................................................29  
Main board....................................................................................................29  
Digital communications .........................................................................................30  
Setting the IP address ......................................................................................30  
Modbus over TCP/IP .........................................................................................30  
Analyzer fault register ................................................................................30  
Remote operation via VNC...............................................................................32  
Spare parts................................................................................................................33  
List of Figures  
Figure 1 Functional pneumatics diagram ........................................................... 12  
Figure 2 Home screen............................................................................................. 17  
Figure 3 Home screen showing active alarm ..................................................... 17  
Figure 4 Home screen showing furnace and ozonator warnings...................18  
Figure 5 Home screen showing main menu choices.........................................18  
Figure 6 Calibration menu, for NOx1 gas............................................................19  
Figure 7 Numeric entry of calibration bottle value using pop-up keypad..... 19  
Figure 8 Diagnostic screen....................................................................................20  
Figure 9 Trend screen.............................................................................................20  
Figure 10 Config screen: Sub-screens for further configuration.................... 21  
Figure 11 Analog output scaling screen ..............................................................22  
Figure 12 Analog output trim screen ...................................................................22  
Figure 13 IP address screen...................................................................................23  
Figure 14 QR code snapshot screen ....................................................................23  
Figure 15 About screen ..........................................................................................24  
Figure 16 Alarm screen, with one alarm cleared and two others active........24  
Figure 17 Config alarm screen ..............................................................................25  
4
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
List of Tables  
Table 1 I/O Analog output pin assignments (typical).......................................14  
Table 2 Expected flows and temperatures during normal analyzer  
operation.................................................................................................... 15  
Table 3 Analyzer fault register bit assignments................................................ 31  
Table 4 Modbus register map............................................................................... 31  
Table 5 Analyzer spare parts list...........................................................................33  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
5
Important precautions  
Use dry, oil-free instrument air only  
Caution: This instrument will be damaged if used with in-  
strument air that is not completely dry and oil-  
free. Ensure instrument air has 40 °C dew point  
and has been filtered to remove all oil and particu-  
lates.  
Safety notice  
This instrument operates from potentially lethal line voltage. In addition,  
some internal components operate at high temperature and can cause seri-  
ous burns. Observe all precautions when using this device, and particularly  
be sure that all devices connected to the instrument are safely wired and  
properly grounded. Always disconnect power to the instrument before  
opening the enclosure or servicing.  
Caution: The analyzer should not be operated without the  
cover in place and the cooling fan fully opera-  
tional.  
The exterior surface of the converter furnace and  
tubes will rise to nearly 80 °C if operated without  
the cover in place. Serious burns can result if the  
proper precautions are not taken.  
6
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Model CL3020 specifications  
Performance specifications  
NOx  
O2  
Measurement technology  
Chemiluminescence using  
all solid-state detection  
Amperometric Zirconium  
oxide cell  
Measurement range  
Full scale range  
0 to 1000 ppm  
0 to 25 % O2  
Continuously adjustable  
from 5 to 1000 ppm  
Continuously adjustable  
from 5 to 25 % O2  
Zero noise  
< 0.04 PPM  
< 0.02 % O2  
Zero calibration drift  
Span noise  
Better than ± 0.1 PPM  
< 0.25% of reading  
Better than ± 1% of reading  
Better than ± 0.1 % O2  
< 0.02 %O2  
Span calibration drift  
Linearity error  
Better than ± 0.1 % O2  
< 2% of high calibration  
value across range from  
zero to full scale1  
< 1% of high calibration gas  
value across range from  
zero to full scale1  
Response time  
T95 < 10 seconds  
> 95%  
T95 < 10 seconds  
NO2 converter efficiency  
Features  
Touch-screen interface: All diagnostics and controls are accessible  
through an advanced, full-color 5” touch screen interface.  
One-touch calibration: Once target gas values (e.g. bottle concentra-  
tions) have been entered, span and zero responses may be captured,  
and hence the analyzer calibrated, at the touch of the screen.  
Trend-screen: Gas concentrations are plotted in a chart-recorder style  
trend with user settable scales for in-depth data analysis at a glance.  
Diagnostics and alarms: Critical component temperatures and gas flows  
are measured within the analyzer and reported on the diagnostic screen.  
Target values and alarmable deviations are user-settable. These alarms  
are displayed on the home screen, as well as warning messages for inter-  
nal communication errors or if the converter or ozonator has been disa-  
bled.  
Analog outputs: Each measurement, including dual ranges for NOx, is  
output as either 4-20mA or 0-10V (user selectable). Analog outputs can  
be forced to low (4mA/0V) or high (20mA/10V) for troubleshooting. Fur-  
ther, the gain and offset of these analog outputs may be trimmed within  
1 Provided the calibration value is 80 to 100% of the full scale.  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
7
approximately +/-5% of full scale to compensate for offsets or other is-  
sues with an external the data acquisition system, independent of ana-  
lyzer calibration.  
QR code (2D barcode): Analyzer configuration and operating parame-  
ters, including calibration settings, can be sent via any mobile device  
that has a QR code / barcode scanning app for ease in remote trouble-  
shooting and support.  
Digital communications: MODBUS over TCP/IP included, allowing access  
to gas concentrations, diagnostics, alarms, and other instrument param-  
eters. The analyzer is also equipped with a VNC server that allows full re-  
mote operation from any device with a VNC viewer connected to the net-  
work. Using a locally connected PC or mobile device, factory personnel  
may remotely inspect and diagnose analyzer problems as if they were  
standing in front of the analyzer.  
Mechanical specifications  
EIA 19-inch rack mount enclosure, 11 in. deep, 3 rack units tall  
(5.25 inches).  
Weight: 24 lbs.  
Power: 120VAC, 4 Amps max.  
Sample flow rate: Requires approximately 0.1 SLPM at atmospheric pres-  
sure (e.g. from a vented sample manifold) per NOx measurement channel  
Instrument air: Requires approximately 0.2 SLPM dry, instrument-quality,  
oil-free air at atmospheric pressure (e.g. from a vented manifold), per  
NOx measurement channel  
8
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Overview  
The Model CL3020 CLD NOx analyzer is designed specifically for use in low-  
NOx CEMS applications. NOx and NO (if dual-channel) are measured using  
chemiluminescence and a molybdenum-based NO2 converter for total NOx.  
The Model CL3020 meets or exceeds 40CFR60 and 40CFR75 demands for  
relative accuracy, linearity, and calibration drift in low and ultra-low NOx  
gas-fired turbine applications.  
Caution: The CL3020 is designed to analyze a clean, dry  
sample, as is typical of a conventional extractive  
CEMS. The sample dew point should be less than 5  
C, and without any appreciable particulates or  
other condensable or reactive gases. As with all  
NOx analyzers, care should be taken in SCR appli-  
cations to scrub any residual ammonia from the  
sample to avoid contamination of internal compo-  
nents.  
Measurement configurations  
The Model CL3020 has several different configurations:  
1. Dual-ranging NOx, with or without O2. In this configuration, total NOx is  
measured using a molybdenum-based converter to convert any NO2 in  
the gas stream to NOx. Two separately calibratable NOx ranges are avail-  
able, e.g full scales of. 100 PPM and 25 PPM, in addition to a single range  
output for O2.  
2. Dual NOx (speciating), with or without O2. In this configuration, two inde-  
pendent chemiluminescence cells are used to simultaneously measure  
one converted stream (total NOx) and one unconverted stream (NO). The  
difference of these two is reported as NO2. This configuration may also  
be used with an external NH3 converter to measure total NOx plus NH3 on  
one channel, total NOx on the other channel, with the difference being  
NH3. A variety of analog output options are available in this configura-  
tion.  
3. Enhanced performance for low range NOx, with or without O2. For full  
scales less than 200 PPM, greater sensitivity may be achieved with modi-  
fied flows. Flows given in this manual are for standard configuration;  
consult factory for more information about flows in this configuration.  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
9
Theory of operation  
The Model CL3020 CLD NOx analyzer uses chemiluminescence as the funda-  
mental detection mechanism for NOx measurement and a pumped (ampero-  
metric) zirconium oxide cell for the O2 measurement. A brief description of  
these mechanisms and their implementation in the analyzer are given be-  
low.  
Chemiluminescence measurement of NOx  
Chemiluminescence is defined as a chemical reaction that gives off light. Ni-  
tric oxide (NO) emits infrared light when it reacts with ozone (O3) to form  
NO2. By introducing a sample containing NO into a reaction cell and mixing  
it with ozone, one can measure the amount of light emitted by the ensuing  
reaction and can infer the amount of nitric oxide present in the original  
sample.  
It is important to note that only NO is the only species directly measured.  
NOx, defined as the sum of NO and NO2 in a sample, is measured by first  
converting any NO2 to NO before it enters the measurement cell using a mo-  
lybdenum-based converter. If so equipped, NO only is measured in a second  
chemiluminescence cell from a sample that does not visit the converter.  
Zirconium oxide measurement of O2  
The oxygen measurement makes use of the fact that zirconium oxide con-  
ducts oxygen ions when heated above approximately 600 °C. Platinum elec-  
trodes on the interior and exterior of a zirconium oxide tube provide a cata-  
lytic surface for the exchange of oxygen molecules and oxygen ions. As mol-  
ecules encounter the platinum electrodes, they become ionized and are  
transported through the body of the zirconium oxide. This charge transport  
ultimately establishes an electric potential across the electrodes that is pro-  
portional to the log of the ratio of oxygen concentrations on each side of  
the oxide (The Nernst Equation). Thus, if a reference gas (usually instrument  
air at 20.9 % O2) flows across the inner electrode, the concentration of sam-  
ple gas flowing across the outer electrode can be determined. In a conven-  
tional zirconium-oxide oxygen analyzer, this voltage is exponentiated to de-  
termine the concentration.  
In the Model CL3020, a second zirconium-oxide cell is ganged together to  
pump oxygen into the first cell, which is maintained at a constant voltage.  
The amount of oxygen needed to maintain the primary cell at the operating  
point is a more sensitive measurement of sample concentration, and allows  
for measurement at zero oxygen. This pump signal is carefully measured  
and related back to the sample concentration.  
10  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Pneumatic design  
Although the chemiluminescence technique is extraordinarily sensitive, spe-  
cific to NOx, and inherently linear, there are many subtle effects involving  
pressures, flows, cell geometry, etc. that must be carefully engineered to  
produce a properly functioning analyzer. The pneumatic design of the Model  
CL3020 is shown in figure 1; the flow scheme is substantially different than  
other gas analyzers. To properly operate and service the analyzer, it is im-  
portant that the flow scheme be well-understood.  
Starting on the upper left side of the diagram, instrument air is drawn into  
the analyzer from a manifold vented to atmospheric pressure. Excess air  
flow should be available at this manifold to ensure integrity of instrument  
air to the analyzer. It is important that the air be free of oil and particulates.  
Ozone for the chemiluminescence reaction is produced in the ozonator, and  
is drawn into the cell where it mixes with the sample stream as described  
earlier. A flow-control orifice is embedded in the fitting on the ozone inlet(s)  
of the reaction cell(s), and the instrument vacuum pulls the proper flow  
through the ozonator.  
Two sample channels, if so equipped, may be present on the analyzer. Ex-  
cess sample flow should be available to each to ensure good sample integ-  
rity to the analyzer. A sample pump configured to deliver 1 SLPM under  
slight positive pressure to a 1/4" SwagelokTM tee connected to the analyzer,  
with a vent tube at least three feet long on each branch is an ideal connec-  
tion.  
Note: Under no circumstances should the sample inputs or ozone  
feed air be pressurized.  
For the total NOx channel, the sample first flows through the converter  
where any NO2 is converted to NO, while any NO present is unaltered. The  
furnace temperature is displayed on the front panel of the analyzer. Next,  
this sample flows though the oxygen sensor (if so equipped), and then  
through the sample flow meter. This flow rate should approximately  
70 SCCM. The sample then flows on to the reaction cell where it mixes with  
the ozone stream and the NOx measurement is performed. The reaction cell  
is kept at 40 °C.  
The exhaust port of the reaction cell contains a critical flow orifice, which  
when backed by a sufficiently strong vacuum, controls the total flow drawn  
through the analyzer. Exhaust from the cell is routed through the high tem-  
perature furnace to destroy all the ozone in the exhaust stream to preserve  
the integrity of downstream components. Unlike carbon filters or other de-  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
11  
signs, this ozone destruction technique is very safe and effective. No meas-  
urable ozone is left in the exhaust. The analyzer is normally configured for an  
external pump to provide critical vacuum to pull the exhaust from the reac-  
tion cell. The flow through the analyzer is independent of this vacuum, pro-  
vided it is low enough to meet the conditions for critical flow. The pressure  
on the downstream side of the critical flow orifice (measured under condi-  
tions of full analyzer flow) should be no more than 30% of atmospheric  
pressure.  
If equipped, a second sample channel is present which does not flow  
through the converter, therefore measuring only NO. The difference be-  
tween these two channels represents the NO2 concentration in the original  
sample, or any such other speciation (e.g. NH3) depending upon external  
system configuration.  
12  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Optional 2nd NO channel  
Prepared sample from  
sample conditioning  
system; .25 to 1 slpm,  
slightly above atm  
Prepared sample from  
sample conditioning  
system; .25 to 1 slpm,  
slightly above atm  
Air Inlet  
1/4“ Swagelok  
1 slpm, max  
dry, oil-free  
pressure, dry to less  
than 1% water content  
pressure, dry to less  
than 1% water content  
Vacuum  
must be less  
than 300 Torr  
at 1.2 slpm  
Vent to atmosphere  
(vent tube should be  
at least 3 feet long to  
prevent diffusion)  
Vent to atmosphere  
(vent tube should be  
at least 3 feet long to  
prevent diffusion)  
Vent to  
atmosphere  
Instrument  
Air Inlet  
Exhaust  
Outlet  
NOx/NO2  
Inlet  
NO Inlet  
Sample 2  
Flow  
Meter  
NO2  
Converter  
Furnace  
Ozone  
Flow  
Meter  
Sample orifice  
fitting  
Oxygen  
Sensor  
NOx cell  
(sample 2)  
Exaust orifice  
fitting  
Sample 1  
Flow  
Meter  
Ozone  
Generator  
Exaust orifice  
fitting  
Ozone orifice  
fitting  
NOx cell  
(sample 1)  
Ozone orifice  
fitting  
Sample orifice  
fitting  
Analyzer Enclosure  
Figure 1 Functional pneumatics diagram  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
13  
Analyzer setup and quick start procedure  
For experienced users already acquainted with gas monitoring techniques  
and equipment, the following is a summary of installation and startup steps  
for a typical CEMS installation; consult factory for other applications. Ana-  
lyzer menus and operation are described in more detail in the next section  
of the manual. To ensure the quickest and most reliable startup, please fol-  
low the five steps below in the order shown.  
Connect the analyzer  
1. Connect instrument air and sample to inlets via 1/4” SwagelokTM fit-  
tings.  
a. Instrument air: 1 SLPM of dry, oil-free air, vented to ambient condi-  
tions. The instrument draws approximately 0.2 SLPM of air per  
NOx channel, and a sufficient excess should be supplied to the an-  
alyzer to ensure the analyzer only pulls instrument air, not ambient  
air, from the vented connection (see figure 1).  
b. Sample #1 (NOx): 0.2 SLPM of sample from a sample manifold  
vented to atmospheric pressure. The instrument draws approxi-  
mately 60 SCCM of sample per NOx channel, and a sufficient ex-  
cess should be supplied to the analyzer to ensure the analyzer only  
pulls sample gas, not ambient air from the vented connection (see  
figure 1).  
c. Sample #2 (NO), if equipped: 0.2 SLPM of sample from a sample  
manifold vented to atmospheric pressure. The instrument draws  
approximately 70 SCCM of sample per NOx channel, and a suffi-  
cient excess should be supplied to the analyzer to ensure the ana-  
lyzer only pulls sample gas, not ambient air from the vented con-  
nection (see figure 1)  
2. Connect 1/4” diameter exhaust line (to pump or eductor) to 1/4”  
Swagelok™ exhaust port. The vacuum source should be able of main-  
tain an absolute pressure of less than 200 Torr at 1 SLPM flow. For ex-  
ample, a Thomas/Gardner model 2688VE44 mechanical pump or Air-  
Vac Engineering AVR-038H air driven eductor.  
3. Connect analog output signals via 8-pin Phoenix Contact connector  
(provided) per the pin assignments listed in Table 1, and/or Ethernet  
connection if using digital communications.  
Warning: This instrument is designed for use with 120V AC  
input power only. Serious equipment damage  
and/or injury will occur if it is connected to im-  
proper power.  
14  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Analog Outputs: The CL3020 analyzer has four analog outputs, assignable  
to various measurements, depending upon configuration. Below are analog  
output assignments typical of a dual-range non-speciating NOx-O2 analyzer.  
Refer to analog output section of this manual for more information.  
NOx output (primary range):  
4-20 mA or 0-10V 0 to full-scale ppm  
NOx output (secondary range):  
4-20 mA or 0-10V 0 to full-scale ppm  
O2 output:  
Pin 1: low  
Pin 2: high  
Pin 3: low  
Pin 4: high  
Pin 5: low  
Pin 6: high  
Pin 7: low  
Pin 8: high  
4-20 mA or 0-10V 0 to full-scale % O2  
Reserved  
Table 1 I/O Analog output pin assignments (typical)  
Apply power to the analyzer and check diagnostics  
Caution: The analyzer should not be operated without the  
cover in place. The exterior surface of the con-  
verter furnace and tubes will rise to nearly 80°C if  
operated without the cover in place. Serious burns  
can result if the proper precautions are not taken.  
1. Apply power by connecting the instrument power cord (provided).  
Verify that the fan is operating by feeling for air flow at the back of  
the instrument. If inadequate flow is suspected, shut down power.  
Caution: The analyzer should not be operated without the  
cover in place and the cooling fan fully functional.  
Care should be taken to avoid blocking the air  
vents in the side panel. Standard EIA rack mount-  
ing should provide enough space for adequate  
cooling.  
2. After approximately one minute the touchscreen will complete its  
startup cycle and be at the home screen. Navigate to the diagnostic  
screen and verify the following as summarized in Table 2:  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
15  
a. Verify that the sample-flow (on the diagnostic screen) indicates  
approximately 70 SCCM per channel and the ozone-flow (diagnos-  
tic screen) indicates approximately 230 SCCM for single channel  
NOx or 460 SCCM for dual channel configuration. If flows are not  
correct, check pneumatic connections and external system com-  
ponents (e.g. pump).  
b. Verify the furnace temperature is rising. The furnace temperature  
should reach 400°C in about 30 minutes.  
c. Verify the ozonator temperature is rising. The ozonator tempera-  
ture should reach 40°C in about 15 minutes. It may be necessary to  
adjust the ozonator setpoint temperature for operation in unusu-  
ally cool or warm environments. The operating temperature is not  
important, it is only necessary that the ozonator temperature re-  
mains constant.  
d. Verify the NOx reaction cell(s) temperatures is rising. The NOx cell  
temperature(s) should reach 40°C in about 15 minutes.  
Parameter  
Value  
Sample flow  
70 SCCM  
Ozone flow  
230 SCCM single NOx, 460 SCCM dual NOx  
Converter temperature  
Ozonator temperature  
NOx cell #1 temperature  
NOx cell #2 temperature (if equipped)  
400 oC  
50 °C (user settable)  
40 °C  
40 °C  
Table 2 Expected flows and temperatures during normal analyzer opera-  
tion  
Set analog outputs of the analyzer  
From the analog output screen located within the Config menu, set full  
scale to desired values for each output channel, per configuration. Refer to  
analog output section of this manual for more information.  
 
16  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Calibrate the analyzer  
After installation and temperatures have reached their setpoints above, the  
NOx and O2 channels can each be calibrated via the following procedure:  
1. Low calibration:  
a. Flow low calibration gas through the sample handling system and  
analyzer. Dry nitrogen, EPA protocol zero gas, or well-scrubbed in-  
strument air is recommended as a low calibration gas for NOx.  
O2 may be zeroed on NO span gas.  
b. Wait approximately two minutes or until reading settles. It may be  
helpful to monitor the trend screen to evaluate when the reading  
has reached final value.  
c. Enter the value of the low calibration gas, typically 0.  
d. Press “Low Capture” soft button on the calibration screen.  
2. High calibration:  
a. Flow high calibration gas through the sample handling system and  
analyzer.  
b. Wait approximately two minutes or until reading settles. It may be  
helpful to monitor the trend screen to evaluate when the reading  
has reached final value.  
c. Enter the value of the high calibration gas, typically from the re-  
ported bottle value from the calibration gas supplier.  
d. Press “High Capture” soft button on the calibration screen.  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
17  
Display screens and details of operation  
Home screen, warnings, and alarms  
The analyzer home screen analyzer is shown in Figure 2 below. In this view,  
only gas concentrations are displayed. Gas concentrations displayed will  
vary by configuration.  
Figure 2 Home screen  
If active alarms are present (any diagnostic variable out of range, see sec-  
tion on Alarm screen), a red triangle with an exclamation point will appear in  
the lower right-hand corner of the display as shown in Figure 3. Touching  
this icon will bring up the Alarm screen as described later.  
Figure 3 Home screen showing active alarm  
If there are active warnings, these descriptors will be annunciated in a ban-  
ner at the bottom of the screen as shown in Figure 4. Possible warnings in-  
clude:  
Communications error between display and main analyzer board,  
warning the display may not be updating thus readings or statuses  
may be invalid.  
Ozonator and/or furnace disabled (see Configuration screen)  
Analog outputs forced high or low (see Analog Output Trim screen)  
 
18  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Figure 4 Home screen showing furnace and ozonator warnings  
Analyzer menus may be accessed by touching the screen anywhere, causing  
the menu bar to become available at the bottom of the screen for about five  
seconds as shown in Figure 5. When the menu is visible, any of the main  
screens (Calibrate, Diagnostics, Trend, Config, of Alarm) may be selected.  
Note: The menu bar will appear from any of the main screens by touching  
the screen anywhere there is not an active input box or button, enabling to  
return to the Home screen from any other main screen.  
Figure 5 Home screen showing main menu choices  
Calibration menu  
Figure 6 shows an example calibration screen, in this case NOx1. Each gas  
has its own calibration screen selectable by touching the corresponding rec-  
tangle from the column of choices on the right of the screen.  
In this screen the raw value displayed corresponds to the percentage of ana-  
log input from the sensor. When troubleshooting, the raw value is often  
more important to examine than the calculated concentration. This is be-  
cause the calculated value may be corrupted by erroneous calibration, but  
the raw value represents the underlying sensor response. The gain is the cal-  
culated correspondence between gas concentration and raw value captured  
during calibration, as described below.  
   
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
19  
The analyzer is calibrated by assigning sensor responses to gas concentra-  
tions. Any two points may be used for calibration, but typically zero and  
span are chosen. Bottle values for calibration may be entered by selecting  
the appropriate field and entering numeric values form the pop-up touch  
keypad, as shown in Figure 7. This same keypad is available for all numeric  
entries throughout the interface.  
Stored raw values corresponding to gas bottle values (or known process val-  
ues) may be “captured” by pressing the high or low capture buttons when  
high or low concentration gas, respectively, is flowed to the analyzer. These  
response values may be entered manually by entering raw values in these  
fields using a numeric pop-up touch keypad.  
Figure 6 Calibration menu, for NOx1 gas  
Figure 7 Numeric entry of calibration bottle value using pop-up keypad  
Diagnostic screen  
Table 2 shows typical values for expected flows and temperatures during  
normal operation. These diagnostics are displayed in the main diagnostic  
screen, and shown in Figure 8. Refer to the Alarms screen for setting limits  
on these parameters, and the troubleshooting section of this manual of if  
these values are out of range.  
   
20  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
The 32-bit Fault Register is defined in more detail in the digital communica-  
tions section. Bit 0 is on the far right, bit 15 is on the far left. Bits are one  
(TRUE) if in alarm condition, zero if not alarmed.  
Figure 8 Diagnostic screen  
Trend screen  
The trend screen is shown in Figure 9. Note that the scale for Oxygen (on  
the left) and NOx reading(s) on the right is adjustable by selecting the top  
and bottom numbers and rescaling with the pop-up keypad. The trend  
screen displays ten minutes of data, refreshing once a second.  
Figure 9 Trend screen  
Config screen  
The configuration screen is shown in Figure 10.  
Furnace power and ozonator output may be toggled by tapping the appro-  
priate buttons. Ozonator temperature will be maintained if disabled, only  
the ozonator discharge itself is disabled. Warnings messages will be dis-  
played on other screens when either of these components is disabled.  
   
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
21  
The furnace temperature setpoint may be adjusted from this screen. During  
steady state operation at room temperature, the power duty cycle should  
be approximately 35%.  
Averaging time for signals may be set to any value between 3 and 60 sec-  
onds. 15 seconds is the default,  
Sub-screens for Analog outputs, IP address, QRC, and About may be ac-  
cessed from this screen and are described below.  
Figure 10 Config screen: Sub-screens for further configuration  
Analog output scaling screen  
The analog output screen is shown in Figure 11, in the case of dual-ranging  
NOx with O2. Other gas configurations are similar and self-explanatory. Con-  
centrations corresponding to low and high analog outputs may be set using  
the numeric pop-up keypad as shown in Figure 7.  
Current (4-20 mA) or voltage (0-10V) output is selectable from the Type  
menu box in the upper right-hand corner. Note the voltage output is in-  
tended for use only with high impedance (>1000 kOhm) devices.  
Actual live outputs are shown for reference  
Note the “gear” icon to the left of each channel. Selecting this icon brings up  
the Analog output trim screen for each channel, as described below.  
 
22  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Figure 11 Analog output scaling screen  
Analog output trim screen  
Each analog output channel may be trimmed independently from its own  
screen, as shown in Figure 12. Under normal operation, the output is in “live”  
mode, but it may be forced to either the high or low limits by selecting the  
corresponding box. When the output is so forced, the corresponding trim  
field is active and may be used to adjust the actual output up or down to  
make up for any system discrepancies. In this manner, the analog loops and  
any external data acquisition system may be calibrated independent of gas  
concentration readings. This may also be used for troubleshooting external  
connections.  
Note that when any outputs are forced high or low, a warning is displayed  
to the user in the warning banner in the lower portion of any main screen.  
The “Next” and “Return” buttons allow the user to cycle through other ana-  
log channels and/or return to the main analog output screen.  
Figure 12 Analog output trim screen  
   
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
23  
IP address screen  
The IP address screen is available from the Config screen. Either a static or  
dynamic IP address screen may be specified.  
Figure 13 IP address screen  
QR code  
A standard QR code (2D barcode) image, as shown in Figure 14, may be dis-  
played for capture and analysis by any mobile device with barcode scanning  
capability. Scanning with a mobile device will provide a text description of  
the analyzer configuration, status, and current readings that can be sent to  
support personnel for troubleshooting assistance.  
Figure 14 QR code snapshot screen  
 
24  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
About screen  
The About screen, as shown in Figure 15, shows the current software ver-  
sions for the system.  
Figure 15 About screen  
Alarm screen  
The alarm screen is shown in Figure 16, accessible from the home screen by  
touching the red triangle. An active, or historical but uncleared, alarm sets  
the corresponding bit in the fault register, and causes the red alarm icon to  
flash on any main screen. If an alarm condition is still being met (causing the  
alarm), the alarm listing will be highlighted red. If the alarm condition is no  
longer being met (not alarmed), the text will no longer be highlighted in red,  
but still visible in the alarm history. The history can be cleared by tapping  
the “Clear” button. Only alarms that are no longer active may be cleared; if  
an alarm condition is still occurring it will reactivate and not remain cleared.  
To access the sub-menus allowing configuration of each alarm, touch the  
Config Alarms button on this screen.  
Figure 16 Alarm screen, with one alarm cleared and two others active  
   
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
25  
Config alarms screen  
Each alarmable temperature and flow is configurable, as shown in Figure 17,  
in this case for the furnace temperature. While factory defaults settings  
should normally not require modification, allowing these to be configurable  
enables the user to operate the analyzer in custom configurations for  
unique applications or for testing purposes, while still having meaningful  
alarm conditions. A target value and deviation from the target may be set in  
the corresponding boxes. Note in the case of a control loop, like furnace  
temperature, this is only the target for the alarm, not the control setpoint  
temperature.  
Figure 17 Config alarm screen  
 
26  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Troubleshooting  
Below is troubleshooting information on the analyzer, as well as information  
on many of the sub-assemblies within the unit. Before troubleshooting the  
analyzer, read through the diagnostic and operating parameters explana-  
tions to help narrow down any problems.  
Warning: Potentially lethal line voltage, lethal high voltage  
(within the ozone generator), and dangerously hot  
tubing and subassemblies are present within the  
analyzer. Service within the analyzer should be  
performed only by qualified, trained personnel,  
and only after the unit has been unplugged and al-  
lowed to cool for at least one hour with the cover  
off.  
Diagnostics and operating parameters  
The diagnostics on the analyzer display much instrument status infor-  
mation (see diagnostics screen). From these readings, it is usually possible  
to narrow down the source of any problems:  
Warnings and alarms  
The Home screen of the display will display the presence of any warnings or  
alarms. Carefully check this information to troubleshoot any problems, for  
example if NOx is not responding it may be the case that the ozonator has  
been left disabled.  
Sample and ozone flow  
Because these two flow rates are intimately related, it is important to con-  
sider them together. Sample flow is the amount of flow drawn through the  
sample inlet through the analyzer. Ozone flow is the amount of feed air  
pulled through the ozonator.  
A critical flow orifice located in the exhaust fitting of the NOx cell deter-  
mines the total flow (sample and ozone flow combined). This total flow rate  
should be approximately 230 SCCM per channel.  
One may determine several things by inspecting the NOx sample and ozone  
flow meters:  
If the both flows are correct, it is very unlikely that there are pneumatic  
problems within the analyzer.  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
27  
If the both flow are low, there may leaks in the analyzer, the critical flow  
exhaust orifice may blocked, the sample inlet may be blocked, the ex-  
haust may be blocked or not vented properly, or the vacuum at the rear of  
the analyzer may be inadequate.  
If the ozone flow is too low, the ozone air feed orifice may be blocked,  
there may be a leak in the ozone supply tubing, or the compressed air  
supply to the instrument may be faulty.  
If the sample flow is too low, the sample orifice may be blocked, there  
may be a leak in the sample tubing, or the sample supply to to the instru-  
ment may be blocked  
If either or both flows are too high, check for pressurization of sample or  
instrument air (inadequate or blocked vents), or other blockages. For ex-  
ample, if the ozone flow is restricted, one indication may be higher than  
normal sample flow, and vice versa.  
Converter temperature  
The converter furnace temperature is maintained at 400 °C. The tempera-  
ture should not vary by more than two degrees. If it is not at the proper  
value, the furnace heater, thermocouple, or relay could be at fault.  
Gas concentration readings  
The O2 value should be stable to approximately 0.02% (absolute). Instability  
or inaccurate readings could be caused by leaks or a faulty O2 sensor mod-  
ule. The NOx reading should be stable to better than 0.25% or reading, or  
0.02 PPM, whichever is greater. Instability or inaccurate readings could be  
caused be leaks, a faulty ozonator, or a faulty NOx sensor or sensor board.  
System checks  
Component failures within the analyzer are relatively uncommon. Most per-  
formance or reliability problems are due to improper system connections,  
leaks, faulty electrical connections, or improper configuration, in that order.  
Before opening the analyzer enclosure, double-check that the external con-  
nections and supply of conditioned sample gas, compressed air, and line  
voltage are correct. The following may aid in troubleshooting if the analyzer  
is not functioning properly and no faults are apparent form the diagnostic  
information.  
Electrical connections and fuse  
Be sure all connectors on the main board are firmly seated and that all wires  
within these connectors are firmly attached. There is a fuse holder on the  
main power inlet of the analyzer, serviceable from the analyzer back panel.  
28  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Leak checking  
Leaks may cause many non-obvious problems, a leak check is highly recom-  
mended when servicing or troubleshooting gas analyzers for any reason.  
Leak checking should be done under vacuum, never pressure with the fol-  
lowing procedure:  
1. Seal the sample and instrument air inlet(s) on the back of the analyzer.  
2. Connect a vacuum pump to the exhaust port, with an isolation valve in  
between the pump and the analyzer, and a vacuum gauge on the ana-  
lyzer side of the isolation valve.  
3. Open the isolation valve, pump the analyzer down, and record the vac-  
uum reading.  
4. Close the isolation valve. The isolated analyzer pressure should not in-  
crease more than 2 Torr/second.  
If leaks are found, check all fittings for tightness, and locate the leak by pro-  
gressively isolating parts of the analyzer. In general, if a leak is present it will  
be very noticeable, not subtle. A slight apparent leak within the ozonator is  
acceptable; a metered sweep of the ozonator assembly is designed to mini-  
mize stray ozone within the analyzer enclosure.  
Component information  
Information on service replacement of many of the analyzer components is  
given below. Only proceed to investigation of components after thoroughly  
checking pneumatic and electrical connections both inside and external to  
the analyzer as noted above. “Part-swapping” without clear indication of  
component failure is not a recommended troubleshooting or repair tech-  
nique.  
Oxygen sensor module  
This is serviced as a unit. If approximately 9VDC is present on pins 1 and 2 of  
the five-pin connector, and oxygen readings are unstable, nonlinear, or inac-  
curate, replace this module. If 9VDC is not present, examine connections.  
Ozonator  
Lethal voltage is present with the ozonator: Do not attempt to service the  
ozone generator, and never open its enclosure or allow tools to come near it  
when the analyzer is energized. The ozonator hums audibly when operating,  
if power is being supplied to this unit (24VDC) and there is no hum, the  
ozonator should be replaced.  
NOx reaction cell  
The NOx reaction cell has several fittings and the NOx sensor mounted on it.  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
29  
The exhaust fitting on the back of the cell houses the critical flow orifice. It  
may be changed by disconnecting the tube leading into the furnace and un-  
screwing the fitting. The orifice and fitting are replaced as an assembly.  
If the readings have dropped, or if the analyzer has been subjected to am-  
monia, it may be necessary to clean the window on the sensor:  
1. Turn off power to the analyzer and remove vacuum from the exhaust port.  
2. Remove the two screws holding the detector assembly to the reaction  
cell and remove the detector assembly.  
3. Wipe off the optical window of the sensor through this port with deion-  
ized water or rubbing alcohol to remove any film or deposits. Do not use  
harsh solvents or abrasive materials.  
4. Reinstall the detector assembly.  
5. Energize analyzer, and apply vacuum. Leak check if readings unstable.  
Converter  
The converter has a cartridge heater and a thermocouple. The heater ele-  
ment and thermocouple connect to the main board near the back of the an-  
alyzer. The converter is typically replaced as an assembly, although the ther-  
mocouple and heater may be field replaced by experienced service person-  
nel. The heater assembly should be replaced if either side of the heater  
shows less than 1 MΩ resistance to the shell or if the heater resistance is not  
approximately 70 Ω.  
The power control relay for the heater is socketed near the rear of the board  
and may require replacement if the converter power supply has been short-  
circuited. The molybdenum charge may be replaced, consult factory for de-  
tails.  
Main board  
If the main board must be replaced, the analyzer should be de-energized  
and all connections removed before attempting to remove the board.  
30  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Digital communications  
The model CL3020 analyzer can act as a MODBUS server (slave) and may  
also be operated remotely via any standard PC or mobile VNC viewer. Addi-  
tionally, a QR-code, i.e. 2D barcode, may be scanned by a mobile device to  
capture a complete snapshot of instrument status that can easily be sent to  
remote troubleshooting personnel.  
Setting the IP address  
The IP address for the analyzer is normally set to a static value of 192.168.1.50,  
and can be changed by tapping the IP Address button on the Config Screen.  
Either a static or dynamically assigned (DHCP) address can be specified.  
Modbus over TCP/IP  
The CL3020 can serve as a MODBUS slave over TCP/IP. The MODBUS regis-  
ter map is shown in Table 4.  
Analyzer fault register  
The 32-bit analyzer fault register reflects the alarm status of the analyzer,  
whereas any non-zero value represents an alarm condition. The fault regis-  
ter bit assignments are listed in Table 3. The lowest bit (0) is set to true if  
any alarm is present allowing for quick reading of a single bit for overall  
alarm status. The fault register is also available in float data-type represen-  
tation of the 32-bit value to allow it to be read along with many other input  
registers using a single read command.  
Bit  
0
Parameter  
True (1) if  
Instrument Fault  
Reserved  
Any alarm  
1…5  
6
O2 Heater  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
7
Sample Flow 1  
Sample Flow 2  
Ozonator Flow  
NOx1 Temp  
8
9
10  
11  
NOx2 Temp  
12  
13  
14  
15  
16…31  
Ozonator Temp  
Furnace Temp  
Analyzer Temp  
Comms Error  
Reserved  
Modbus server version 1.00100  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
31  
Table 3 Analyzer fault register bit assignments  
Parameter  
Register Data Notes  
Num*  
Type  
Int32  
Int32  
Int32  
Int32  
Modbus Slave Version  
30001  
System Board Firmware Version 30003  
Display Software Version  
Status (integer)  
30005  
30007  
30021  
30023  
30025  
30027  
30029  
30031  
30033  
30035  
30037  
30039  
30041  
30043  
30045  
30047  
30049  
30051  
30053  
30055  
40101  
40103  
40105  
40107  
40109  
40111  
Status (float)  
Float Float of (Int32 Status) to allow all float reads  
Float Live concentration in gas-units  
Float Live concentration in gas-units  
Float Live concentration in gas-units  
Concentration Gas NOx 1  
Concentration Gas NOx 2  
Concentration Gas O2  
Reserved  
Raw Response Gas NOx 1  
Raw Response Gas NOx 2  
Raw Response Gas O2  
Reserved  
Float Live signal response in % of available input  
Float Live signal response in % of available input  
Float Live signal response in % of available input  
Temperature Analyzer  
Temperature NOx 1 Cell  
Temperature NOx2 Cell  
Temperature Ozonator  
Temperature Furnace  
Heater Power Furnace  
Flow Sample 1  
Float degrees C  
Float degrees C  
Float degrees C  
Float degrees C  
Float degrees C  
Float Live furnace power in %  
Float Live sample 1 flow in cm3  
Float Live sample 2 flow in cm3  
Float Live ozone flow in cm3  
Flow Sample 2  
Flow Ozone  
Stored Span Response NOx 1  
Stored Zero Response NOx 1  
Stored Span Response NOx 2  
Stored Zero Response NOx 2  
Stored Span Response O2  
Stored Zero Response O2  
Float Stored response in % of available input  
Float Stored response in % of available input  
Float Stored response in % of available input  
Float Stored response in % of available input  
Float Stored response in % of available input  
Float Stored response in % of available input  
Float Bottle concentration in gas-units  
Float Bottle concentration in gas-units  
Float Bottle concentration in gas-units  
Float Bottle concentration in gas-units  
Float Bottle concentration in gas-units  
Float Bottle concentration in gas-units  
Span Bottle Concentration Gas 1 40113  
Zero Bottle Concentration Gas 1 40115  
Span Bottle Concentration Gas 2 40117  
Zero Bottle Concentration Gas 2 40119  
Span Bottle Concentration Gas 3 40121  
Zero Bottle Concentration Gas 3 40123  
Converter Furnace Enabled  
Ozonator Enabled  
1
Bool  
Bool  
True (1) enables furnace power  
True (1) enables ozonator power  
2
* Register numbers 30xxx are input registers, 40xxx are holding registers,  
x are output coils. Modbus server version 1.00100  
Table 4 Modbus register map  
   
32  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
Remote operation via VNC  
It is possible to operate the CL3020 remotely by any VNC viewer connected  
via TCP/IP. Every display and every action that is available from the analyzer  
display is duplicated on the remote device, with a mouse click or screen  
touch on the remote device serving as the same input as front screen touch  
commands.  
Any VNC viewer may be used, once the IP address of the analyzer is known.  
The VNC connection password is CL3020 (case sensitive) and cannot be  
changed. TCP port 5900 is used for the VNC connection and must not be  
blocked by any routers or firewalls between the analyzer and the remote de-  
vice.  
CL3020 OPERATING INSTRUCTION | OI/CL3020-EN REV. A  
33  
Spare parts  
Part description  
Exhaust orifice fitting  
Sample orifice fitting  
Ozone orifice fitting  
NOx detector O-ring  
Fuse  
Part number  
1000-1020  
1000-1030  
1000-1031  
1000-1032  
1000-1033  
1000-1034  
1000-1035  
1000-1036  
1000-1037  
1000-1038  
1000-1039  
1000-1040  
1000-1041  
1000-1042  
1000-1043  
1000-1044  
1000-1045  
1000-1046  
1000-1047  
1000-1048  
1000-1049  
Recommended on-site quantity  
2
2
2
2
2
Oxygen sensor  
1 (if configured)  
Optional  
None  
None  
None  
None  
None  
None  
None  
None  
None  
None  
None  
None  
None  
None  
Ozone generator  
NO2 furnace assembly  
Furnace media recharge  
NOx Detector  
Flowmeter  
Furnace relay  
Furnace heater  
Furnace thermocouple  
Furnace sample fittings  
Display  
Electronics board  
Fan  
Sample tubing, 5 ft  
Exhaust tubing, 2 ft  
Tubing tee  
Table 5 Analyzer spare parts list  

Nokia 1110i User Manual
Motorola Stature i9 User Manual
Lindy 51188 User Manual
LG Electronics MM 535 User Manual
Hitachi Deskstar IC35L090AVV2070 User Manual
Gemini CDJ 02 User Manual
GE 106804 User Manual
Fujitsu MPE3XXXAT User Manual
Clarion ProAudio DXZ535 User Manual
Archos TV Plus 500973 User Manual