Thermo Serv Weather Radio PHP601 User Manual

high precision thermometer  
PHP 601  
Resolution 0.001°C  
Main applications include:  
Temperature measurements using RTDs and/ or  
thermocouples  
Absolute or differential measurements over two  
channels.  
Differential thermal analysis.  
Checking the temperature stability of furnaces or  
baths.  
Accuracy 0.005°C  
Self-contained  
RTDs and thermocouples  
Dual inputs  
RS 232 or IEEE 488 interfaces  
Monitoring (alarm) and temperature recording.  
functions  
Resistance measurement (1) ..................................................................................  
The PHP 601 is a high accuracy, dual  
input thermometer suitable for use with  
RTDs (e.g.: Pt 100) and thermocouples.  
The PHP 601 is capable of storing sensor  
characteristics and coefficients and can  
scan up to 12 sensors using the model  
SHP 101 Scanner.  
Its main applications include the calibra-  
tion of temperature sensors. By using our  
CAL'EXPERT software to drive the  
Accuracy (2)  
All ranges  
Stability 24 hours  
90 days  
1 year  
25 to 3 200 W  
0.0005%Cal  
0.0015% + 0.0005%  
0.0030% + 0.0005%  
(1) 4 wire configuration, all currents. For the 3 wire configuration, add 1mW.  
(2) Accuracy is expressed in ± (% reading + % range) at 23°C ± 1°C.  
PHP 601 thermometer, scanner and a ther-  
mal device such as temperature bath, dry  
block calibrator or oven, a complete cali-  
bration procedure can be performed,  
reports generated and certificates printed  
for individual sensors.  
DC voltage measurement .......................................................................................  
Accuracy (1)  
Measurement  
90 days  
1 year  
Range  
range Resolution Stability 24 hours  
100 mV - 50 to + 117.5 mV  
0.2 µV 0.001% + 0.4 µV 0.004% + 0.6 µV 0.008% + 0.8 µV  
(1) Accuracy is expressed in ± (% reading + nV) at 23°C ± 1°C.  
Temperature coefficient £ 10% of accuracy/ °C over 90 days.  
1
 
Temperature measurement with ther-  
mocouples ...........................................  
Types of sensors:  
- Standardized according to IEC  
Publication 580-1/ 1995 (thermocouples  
K, T, J, E, R, S, B and N).  
Thermocouple C according to Hoskins  
curve.  
Thermocouple Platinel (Pl) according to  
Engelhard curve.  
- Special thermocouple Molybde-  
num/ Nickel Molybdenum (Mo).  
- Polynomial equations defined per point  
(25 points max.).  
True differential measurements.  
Programmable internal or external refer-  
ence junction compensation.  
For sensors not complying to the  
International Temperature Scale (ITS 90)  
correction of the deviation ITS 90-IPTS 68,  
1993 is applied.  
Thermocouples U and L according to  
DIN 43710.  
Accuracy (1)  
Thermocouple  
K
Measurement range  
Resolution  
Stability 24 hours  
90 days  
1 year  
- 250 to - 220°C  
- 220 to - 100°C  
- 100 to + 1 350°C  
0,05°C  
0,02°C  
0,01°C  
0,2°C  
0,03°C  
0,0015 % + 0,01°C  
0,25°C  
0,06°C  
0,005 % + 0,015°C  
0,5°C  
0,1°C  
0,01 % + 0,02°C  
- 250 to - 220°C  
- 220 to - 90°C  
- 90 to + 400°C  
0,05°C  
0,02°C  
0,01°C  
0,1°C  
0,03°C  
0,015°C  
0,15°C  
0,06°C  
0,025°C  
0,3°C  
0,1°C  
0,04°C  
T
J
- 210 to - 100°C  
- 100 to + 1 200°C  
0,02°C  
0,01°C  
0,03°C  
0,001 %+ 0,01°C  
0,05°C  
0,004 % + 0,015°C  
0,1°C  
0,04°C  
- 250 to - 180°C  
- 180 to - 100°C  
- 100 to + 980°C  
0,05°C  
0,02°C  
0,01°C  
0,1°C  
0,02°C  
0,001 %+ 0,01°C  
0,15°C  
0,04°C  
0,004 % + 0,01°C  
0,3°C  
0,06°C  
0,008 % + 0,02°C  
E
- 250 to - 175°C  
- 175 to - 100°C  
- 100 to + 1 300°C  
0,05°C  
0,02°C  
0,01°C  
0,2°C  
0,04°C  
0,02°C  
0,25°C  
0,06°C  
0,004 % + 0,025°C  
0,5°C  
0,1°C  
0,008 % + 0,03°C  
N
- 50 to + 400°C  
+ 400 to + 1 768°C  
0,05°C  
0,02°C  
0,1°C  
0,05°C  
0,15°C  
0,004 % + 0,06°C  
0,2°C  
0,01 % + 0,1°C  
S
R
B
U
L
- 50 to + 400°C  
+ 400 to + 1 768°C  
0,05°C  
0,02°C  
0,1°C  
0,05°C  
0,15°C  
0,004 % + 0,06°C  
0,2°C  
0,008 % + 0,08°C  
+ 100 to + 400°C  
+ 400 to + 1 820°C  
0,2°C  
0,05°C  
0,4°C  
0,1°C  
0,6°C  
0,15°C  
1,0°C  
0,2°C  
- 200 to - 70°C  
- 70 to + 600°C  
0,02°C  
0,01°C  
0,03°C  
0,01°C  
0,05°C  
0,004 % + 0,015°C  
0,08°C  
0,008 % + 0,02°C  
- 200 to - 70°C  
- 70 to + 900°C  
0,02°C  
0,01°C  
0,02°C  
0,01°C  
0,04°C  
0,004 % + 0,015°C  
0,08°C  
0,008 % + 0,02°C  
C
Pl  
- 20 to + 2 310°C  
- 100 to + 1 400°C  
0 to + 1 375°C  
0,02°C  
0,02°C  
0,05°C  
0,02 % + 0,03°C  
0,0015 % + 0,01°C  
0,05°C  
0,008 % + 0,05°C  
0,005 % + 0,02°C  
0,05°C  
0,015 % + 0,07°C  
0,01 % + 0,03°C  
0,10°C  
Mo  
(1) Accuracy is given for a reference junction at 0°C in ± (% reading + number of units) or in ± (number of units) at 23°C ± 1°C.  
The uncertainty due to the internal reference junction is ± 0.15°C.  
Temperature coefficient of the internal reference junction: ± 0.015°C/ °C.  
general specifications  
Presentation........................................  
ABS bench-type unit with tilt bail/ handle.  
Dimensions: 225 x 88 x 310 mm.  
Three languages available for the menus  
and on-line help (English, French and  
German).  
Connection using 4-mm plugs and LEMO  
sockets.  
Weight: 2 to 3 kg depending on options.  
Mains supply and optional battery pack.  
LCD graphic display with backlighting.  
scanner SHP 101  
(See separate leaflet).  
to mix both types of inputs.  
Outstanding repeatability between chan-  
nels: Dt £ 0.02°C.  
Switching the channels may be performed  
either automatically by the PHP, or manu-  
ally through the keypad on the front  
panel.  
This is a bench-type scanner dedicated to  
low level signals and intended to switch  
12 (2 wire) inputs (thermocouples) or 6  
(4 wire) inputs (RTDs) with the possibility  
Easy sensor connection by use of screw  
terminal blocks.  
3
 
calibration software LCL30  
(See separate leaflet)  
This software enables the user to define  
automatic calibration procedures.  
It provides control of generated or simulat-  
ed temperatures and will measure the dif-  
ference between the standard thermometer  
and the sensor being calibrated  
This software allows:  
• to program the calibration procedure.  
• to select a standard sensor from a list  
which will be used as reference; this list is  
updated by the user.  
• to control the thermal reference source.  
• to collect the measurements.  
• to print and save calibration reports for  
all the sensors.  
It also enables the user to document and  
recall the history of calibrated sensors.  
accessories  
• Standard sensor AN 5847 and "work-  
ing" standard sensor AN 5848 are  
offered with various connections: plug,  
DIN socket or LEMO socket.  
LEMO-DIN adapter for connection to the  
probe equipped with DIN socket  
(PEM 40316).  
• T 1200 sensor, standard thermocouple  
connectable by using plugs.  
• Cable, 2-meter length (ACL 4603) with  
LEMO socket at one end and RTD free  
sensor connection at the other end.  
ordering instructions  
High accuracy thermometer  
PHP 601-1  
High accuracy thermometer with battery pack and charger PHP 601-2  
High accuracy thermometer with IEEE 488.2 interface  
High accuracy thermometer with battery pack, charger  
and IEEE 488.2 interface  
PHP 601-3  
PHP 601-4  
Accessories ..................................................................................  
Carrying case  
Right-angle brackets for panel mounting  
Rack mounting kit  
AN 6901  
AN 5883  
AN 5884  
RS 232 cable, 9-pin male/ 9-pin female  
IEEE 488 cable, 2-meter  
LEMO socket to be wired  
AN 5875  
AN 5836  
ER 48379  
Cable, 2-meter, LEMO/ free  
Pt 100 standard sensor with plugs  
LEMO Pt 100 standard sensor  
Pt 100 reference sensor with plugs  
LEMO Pt 100 reference sensor  
LEMO Pt 100 reference sensor  
Standard thermocouple  
ACL 4603  
AN 5847B  
AN 5847C  
AN 5848-20000A  
AN 5848-30000A  
AN 5848-30001A  
T 1200  
LEMO-DIN adapter  
Automatic calibration software  
PEM 40316  
LCL 30  
B.P. 182 - F 91006 EVRY Cedex  
Tel. 33 1 69 36 50 60  
Fax 33 1 60 77 82 97  
Email: export@aoip.com  
4
Specifications are subject to modification without prior notice  
 

Teac Stereo System CDX10I User Manual
Technics Stereo System SL BD22 User Manual
Timex Watch 238 095001 User Manual
Toastmaster Blender RRFP1A T TM User Manual
TP Link Network Router TL MR3046 User Manual
TP Link Switch T1500 28PCT User Manual
Triarch Indoor Furnishings 75307 10 User Manual
Tripp Lite Network Cables B090 A9F X User Manual
Vivitek DVD VCR Combo DVD Combo User Manual
Weed Eater Blower 545117525 User Manual