Princeton Marine Sanitation System FA 448 User Manual

15 Discovery Way, Acton, MA 01720  
Phone: (978)263-3584, Fax: (978)263-5086  
Web Site: www.piacton.com  
FA-448 SIX POSITION FILTER WHEEL  
OPERATING INSTRUCTIONS  
Rev: 1.597  
 
I. Description  
The FA-448 is a series of filter wheel assemblies which are available with and without installed order  
sorting filters and with and without remote position control. This manual covers the following  
models:  
FA-448  
Six position filter wheel.  
Manual sample indexing.  
Holds 6 samples or filters, each 1 inch diameter.  
Does not include filters.  
FA-448-1  
FA-448-2  
Order sorting filter assembly.  
320nm, 590nm, 665nm, and 715nm order sorting  
filters mounted in the FA-448 filter wheel assembly. *  
Motorized six position filter wheel with stepping motor  
and FA-448-4 controller. Enables computer controlled  
filter indexing through RS-232 port and front panel  
switch. Does not include filters.  
FA-448-3  
FA-448-4  
Motorized order sorting filter assembly. Same as  
FA-448-2, except with order sorting filters installed.  
Requires computer or terminal with RS-232 port for  
external control. *  
Controller retrofit unit. For use with FA-448 and  
FA-448-1 manual filter assemblies to allow computer  
control or front panel switch control.  
* Positions of order sorting filters in FA-448-1 and FA-448-3:  
Position 1: Open  
Position 2: 320nm cut-off filter  
Position 3: 590nm cut-off filter  
Position 4: 665nm cut-off filter  
Position 5: 715nm cut-off filter  
2
 
II. FA-448 Series Filter Assembly Specifications  
• Six position wheel  
• One inch (1" ±0.005) diameter filters, maximum thickness =0 .200"  
Controller specifications  
• Control from computer or terminal through RS-232 port at 9600  
• baud, no parity, 8 data bits, 1 stop bit and 1 start bit.  
• Optional control from IEEE controller  
Device #10*  
• Time required for filter change = 2.5 seconds for adjacent filter positions.  
Default Address - This address is user settable. See Section IV-C of this manual.  
3
 
III. Installation  
A. Mounting to Monochromator  
The filter assembly is 3/4 of an inch thick. Therefore, if using a light source that is designed to  
mount directly on the entrance slit housing and to focus the light on the entrance slit, it is  
recommended that the filter assembly be mounted on the exit slit housing. If it is necessary to  
install or change filters, do so before mounting the filter wheel to the SpectraPro™ (refer to step C).  
The filter assembly is designed to mount directly to the SpectraPro™ slit housings or to be mounted  
between the light source or detector and the slit housing. The filter assembly is attached to the slit  
housing with three screws. The fourth screw is not used with the filter assembly. Assure that the (3)  
8-32 mounting screws extend approximately 0.25 inch (6mm) beyond the filter assembly housing.  
Orient the filter assembly so that the aperture slot is vertical and the motor is away from the  
instrument housing. Mount the filter assembly or light source/detector to the desired slit housing  
with (3) 8-32 screws.  
B. Connecting Control Unit  
Each control unit must be connected to an AC line of 115 vac. There is also a cable with a 15 pin  
connector attached to each control unit which must be connected to the filter wheel assemblies. For  
RS-232 operation of the filter assemblies, each control unit must be connected to separate RS-232  
ports of a computer or terminal. Acton Research Corporation offers the following connecting cables:  
CC-499-1  
CC-499-2  
CC-499-3  
Cable for IBM-PC or XT compatible computers,  
25 pin female connector to 9 pin male connector  
Cable for IBM-PC/AT or compatible computer,  
9 pin female connector to 9 pin male connector  
Cable for computer terminal  
25 pin male connector to 9 pin male connector  
If none of the above cable assemblies are compatible with your computer or terminal, contact Acton  
Research Corporation for a custom cable or one may be constructed using a standard DE-9P  
connector at the filter controller end with the following connections:  
pin #  
description  
1
2
3
4
5
6
7
8
9
open  
RD received data to the computer  
TD  
open  
ground  
open  
RTS  
CTS  
open  
transmitted data from the computer  
connect these two  
pins together  
For the optional IEEE-488 operation, connect a standard IEEE cable between the two control units  
and connect an additional IEEE cable from one control unit to the IEEE controller.  
4
 
C. Installing and Changing Filters  
To install or remove filters or samples, the following procedure is recommended:  
1)  
2)  
3)  
If the filter assembly is mounted to the SpectraPro™, remove it and disconnect the cable, if  
attached.  
Place the filter assembly on a clean surface with the motor down. Refer to  
Figure 1.  
Using the 1/16" allen wrench supplied, remove the eight screws securing the cover plate to  
the filter assembly (refer to Figure 1). While pushing on the end of the shaft with a pencil,  
carefully remove the cover plate by lifting straight off the shaft. Caution: A spacer is located  
on the filter wheel shaft between the filter wheel and cover. If this spacer comes off with the  
cover, place it back onto the shaft. Refer to Figure 2.  
4)  
5)  
The six filter positions are now visible, refer to Figure 2. The filters are held in with split rings.  
Determine the filters to be removed or the location for the filters to be installed. The filter  
positions are marked 1 thru 6 on the wheel. Remove the split rings. Remove or install filters  
as required. Replace the split ring assuring it is pushed firmly against the filters.  
Assure that the spacer is on the filter wheel shaft and attach the cover plate with the eight  
screws. The filter assembly is now ready for mounting to the SpectraPro™.  
5
 
IV. Operation  
A. Remote Operation from Controller Unit  
The filter positions of each filter wheel can be controlled using the thumbwheel switches on the  
controller modules supplied with the filter assemblies. Connect the motor drive cable of the  
controller to the motor connector of the filter wheel for each unit. Connect the control unit to an AC  
line of the voltage listed on the back of the controller. The positions 0 - 6 may then be selected by  
this switch.  
B. Operation with RS-232 Computer or Terminal  
Connect the motor drive of the control unit to the filter wheel and the RS-232 output of the control  
unit to an RS-232 port of a computer or terminal. See section III-B for cable installation instructions.  
The filter wheel with controller responds to commands from the computer or terminal in the form of  
ASCII strings. The following two commands are used for controlling the filter position:  
FILTER  
Controls the filter position and is used as follows:  
3 FILTER <enter> sets filter wheel to position 3. Filter  
positions 1 - 6 may be selected. Note that there must be a  
after the number and the command must be terminated  
with the enter or carriage return key. The filter will go  
to the specified position in the direction which  
requires the shortest move. When the move is  
complete, the controller will respond with OK followed  
by carriage return and line feed (hex AXCII sequence  
20 6F 6B of OA). Also, the default condition is to echo each  
character that is sent to the controller. If no echo is  
desired, the command NO-ECHO will suppress the echo.  
The command ECHO will return the controller to the  
default echo state.  
?FILTER  
FHOME  
Returns the current filter position. Typing ?FILTER <enter>  
will cause the current filter position to be returned to the  
computer and displayed.  
Initializes the filter wheel to the first filter position.  
This command is automatically executed at power-up.  
The filter wheel controller keeps track of filter  
position as long as power is applied. It should not be  
necessary to execute this command under normal  
circumstances, although it is available if desired. This  
command follows the same format as in FILTER above  
for entry and response.  
6
 
C. IEEE -488 Interface Option for Filter Controllers  
The IEEE-488 interface option allows control of ARC Filter Control unit from the IEEE-488 bus. All  
of the commands and status requests described in Section B - Operation with RS-232 Computer or  
Terminal - are available through the IEEE-488 option. The following standard interface functions are  
provided:  
SH1, AH1, T2, L2, SR1, RL1 (NOTE: The local lockout command  
is not accepted but the front panel is automatically locked out when  
in remote), PP0, DC0, C0  
The GPIB device identifier or address may be sent by dip switches on the interface board. The  
switches are accessible on the bottom of the control unit by removing screws to the access plate.  
The factory set device address is 10. The command ?ID will report back the current device address.  
The following table shows the function of each switch position:  
S1-1 on = address 1 decimal, off = address 0  
S1-2 on = address 2 decimal, off = address 0  
S1-3 on = address 4 decimal, off = address 0  
S1-4 on = address 8 decimal, off = address 0  
S1-5 on = address 16 decimal, off = address 0  
S1-6 factory set to on  
S1-7 factory set to off  
S1-8 factory set to off  
S1-9 factory set to on  
S1-10 factory set to on  
The last byte of each command string to the Filter Control unit must be a <CR> (hex OD). The end  
message (EOI line) is accepted but not required. When the SpectraPro™ is a talker, each string  
sent back to the controller ends with a <LF> (hex 0A) with the EOI line set.  
The Filter Control unit will issue a service request based upon the contents of the service request  
mask which is set by the SET-MASK command and read by the ?MASK command. The bits of the  
mask are as follows:  
bit 0 previously sent command is complete  
bit 1 previous command generated an error  
bit 7 previous command generated a response which is  
now ready to be sent  
The default value for the service request mask is ø, therefore, no service request is generated. To  
change the mask to generate a service request on error or response ready, for example, issue the  
command 130 SET-MASK <CR>. This new value is reset to the default value on reset or power-up.  
The Filter Control unit responds to a serial poll with a status byte that uses the same bit pattern as  
the service request mask shown above.  
The following page is a Basic program listing illustrating the use of the filter controllers from a  
7
National Instruments PC-IIA GPIB Controller.  
 
Program  
1 CLEAR , 59000! : IBINIT1=59000! : IBINIT2=1BINIT1+3 : BLOAD “bib.m”,IBINIT1  
2 CALL IBINIT1 (IBFIND, IBTRG, IBCLR, IBPCT, IBSIC, IBLOC, IBPPC, IBBNA, IBONL, IBRSC,  
IB SRE, IBRSV, IBPAD, IBSAD, IBIST, IBDMA, IBEOS, IBTMO, IBEOT, IBRDF, IBWRTF, IBTRAP)  
3 CALL IBINIT2 (IBGTS, IBCAC, IBWAIT, IBPOKE, IBWRT, IBWRTA, IBCMD, IBCMDA, IBRD, IBRDA, IBSTOP, IBRPP,  
IBRSP, IBDIAG, IBXTRC, IBRDI, IBWRTI, IBRDIA, IBWRTIA, IBSTA%, IBERR%, IBCNT %)  
4 GOTO 20  
7 SAVE”DEVICE12",A:END  
20 UDNAME$ ="DEV12"  
30 CALL IBFIND (UDNAME$,UD%)  
40 IF UD% < 0 THEN TOGO 200  
43 INPUT “ENTER 1 TO GET ?FILTER, 2 TO SEND FILTER, 3 TO EXIT: “,Q  
44 IF Q=1 THEN 50  
45 IF Q=2 THEN 400  
46 IF Q=3 THEN 430  
47 GOTO 43  
50 WRT$="?FILTER” +CHR$ (&HD)  
60 CALL IBWRT (UD%,WRT$)  
70 IF IBSTA% < 0 THEN GOTO 300  
113 PRINT “IBSTA%=";IBSTA%  
116 CALL IBRSP (UD%,SPR%)  
117 IF ( (SPR% AND 129) <> 129 ) THEN 116  
125 PRINT “SPR = “;SPR%  
130 RD$=SPACE$(255)  
140 CALL IBRD (UD%,RD$)  
145 IF IBSTA% < 0 THEN GOTO 300  
150 PRINT “ DATARECEIVED ="; LEFT$(RD$,IBCNT%);  
160 PRINT “ STATUS =" ; IBSTA%  
170 GOTO 43  
200 PRINT “ibfind error “: STOP  
300 PRINT “gpib function call error” : STOP  
400 WRT$="2 FILTER” + CHR$ (&HD)  
410 CALL IBWRT (UD%,WRT$)  
420 IF IBSTA% < 0 THEN GOTO 300  
425 GOTO 43  
430 CALL IBLOC (UD%)  
440 IF IBSTA% < 0 THEN GOTO 300  
500 END  
rem  
rem  
rem  
rem  
rem  
rem  
The above program is for use with the National Instruments  
PC-IIA GPIB controller card for controlling filter wheel A.  
to control filter wheel B, change line 20 to read “DEV13"  
instead of “DEV12". The first three lines of the program  
are merged from National Instruments file DECL.BAS and may  
be different for your computer or GPIB controller card.  
rem  
rem  
rem  
rem  
rem  
To operate, load and run as a BASIC program and enter  
1 to request the filter position, 2 to sent the filter  
to position 2 ( this can be changed in line 400 or made  
a variable ) or enter 3 to return the control unit to  
local control mode ( thumbwheel switch ).  
8
 
V. Figures  
Cove r Sc re ws  
(1 o f 8)  
Sha ft  
Ap e rture Slo t  
Cove r Pla te  
De te nt Sc re w fo r  
Ma nua l Op e ra tio n  
Sp lit Ring  
(1 o f 6)  
Sp a c e r  
Filte r o r Sa m p le  
Po sitio n (1 o f 6)  
6 Po sitio n  
Filte r Whe e l  
9
 
VI. Appendix A: Connector W iring  
Filter Wheel Drive Cable  
J1 Filter Wheel  
Pin #  
1
2
3
4
Motor phase 1A  
Motor phase 1C  
Motor phase 1B  
open  
5
6
7
8
Motor phase 2A  
Motor phase 2C  
Motor phase 2B  
open  
9
open  
10  
11  
12  
13  
14  
15  
Interrupter module OUT  
Interrupter module GND  
Interrupter module LED cathode  
Interrupter module LED anode  
Interrupter module +5vdc  
Shield  
NOTES:  
1. J1 connector is Cannon DAMA-15P. Mating connector is Cannon DAMA-15S.  
2. Cable is 24 AWG with overall shield.  
3. Filter Driver Cable is permanently attached to the Filter Control Module.  
10  
 
VII. Wa r r a n t y  
Certification  
Acton Research Corporation (ARC) certifies that this instrument was thoroughly tested and  
inspected and found to meet the specifications furnished by ARC when it was shipped from  
the factory.  
Wa r r a n t y  
Acton Research Corporation (ARC) instruments and accessories are warranted for a period  
of one full year from date of delivery to be free from defects in material and workmanship and  
to conforms to the specifications furnished by ARC. The corporation's obligation under this  
warranty is limited to servicing or adjusting an instrument returned to the factory, prepaid, and  
to repairing or replacing at the factory any part or parts thereof. All purchased items carry the  
original manufacturers guarantee.  
Acton Research Corporation shall not be liable for consequential damages resulting from  
accident, alteration, misuse, improper installation, operation on low or excessive voltages or  
any use in violation of the operating instructions furnished by ARC.  
If any defect appears within the warranty period, the purchaser shall promptly notify ARC. No  
material will be accepted for repair or replacement without prior authorization from ARC.  
Upon such authorization and in accordance with instructions of Acton Research Corporation,  
parts, materials or equipment for which repair or replacement is requested shall be returned  
to ARC for examination, with shipping charges prepaid by the purchaser. Final determination  
as to whether a product or part is actually defective rests with Acton Research Corporation.  
In such cases where necessary repairs are not covered by this warranty, an estimate of repair  
charges will be submitted to the purchaser before servicing the equipment.  
Acton Research Corporation reserves the right to make changes or improvements upon its  
products without imposing any obligations upon itself to install the same upon its products  
previously manufactured.  
This warranty is expressly in lieu of all other obligations or liabilities on the part of ARC, and  
ARC neither assumes, nor authorizes any other person to assume for them, other obligations  
or liability in connection with the sale of equipment manufactured by Acton Research  
Corporation.  
11  
 

Philips Electric Shaver RQ1077 User Manual
Philips MP3 Player Accessories SJM2106 User Manual
Philips Network Card SWV3458W 17 User Manual
Philips VCR VR354 User Manual
Pioneer Home Theater System S DV131 User Manual
Pioneer Speaker TS 171PRS User Manual
PLUS Vision Projector U5 232 User Manual
Poulan Lawn Mower 183050 User Manual
Poulan Snow Blower 416751 User Manual
ProForm Treadmill 831295230 User Manual